ELI Beamlines
High Harmonic Generation (HHG)

The HHG is a kHz source of ultra-short pulses in the EUV to soft X-rays spectral range. Due to the interaction of noble gases atoms with a strong laser field of 10^15 W/cm2, the emitted electrons will be accelerated and then rescattered to parent atoms to generate radiation of short wavelength in attosecond bursts repeating every laser cycle.

Laser-based light source parameters

High Harmonic Generation

Source type
HHG
Fluctuations of the peak photon energy / central wavelenght
Depends on L1 laser fluctuations
Beam shape
Gaussian
Pulse duration FWHM
20 [fs]
Polarisation
Circular, Elliptical
Other polarisation
linear (later - circular, elliptical)
Pulse repetition rate
1000 [Hz]
Note
All HHG data are shown in a separate document called "HHG-data".

- ELips end-station: No possibility to connect external end station for ELips. Controle Software type: Open source. Data output format: txt. Data output type: cvs. Softwares for data analysis: external Transmission and Ellipsometry are the techniques for the ELips chamber. ELips chamber will be available for the users beginning of 2019.

- MAC chamber: No possibility to connect external end station. MAC will be available on September 2018.
Additional Lightsources

HHG

Beam shape
Gaussian
Optics

HHG-optics

Angle Of Incidence Of Light On Sample
Straight beam
Diagnostics
Spectrometer, X-ray camera, XUV photodiode
Monochromator
Yes
Endstations or Setup

ELips

Detectors Available
Andor iKon-L SO
Endstation Operative
Yes

Sample

Sample Type
Other: Solid state wafers
Mounting Type
Glue by silver paint
Required Sample Size
X = 10000 [um], Y = 10000 [um], Z = 500 [um]

Manipulator or Sample stage

Degrees Of Freedom
5
Cradles
1

Sample Environment

Pressure (min)
1 * 10-7 [mbar]
Temperature
20 - 450 [K]

Magnetic Fields

DC Fields
Magnetic head -> 1.5 T at the sample

MAC chamber

Microscopes
In-line microscope for imaging HHG focal spot.
Base Pressure
1 * 10-8 [mbar]
Detectors Available
MCPs and Cameras

Sample

Sample Type
Other: MAC chamber is mainly for samples in gas phase – aerosols, clusters. Sample delivery systems: • GDVN injector • Flat sheet jet • Cluster source
Mounting Type
It is possible to use a solid sample. A motorised xyzzy and rotation manipulator is available.

Manipulator or Sample stage

Positioning Precision
X = 150 [nm], Y = 150 [nm], Z = 75 [nm]
Range Of Movement
X = 100 [mm], Y = 100 [mm], Z = 50 [mm]

Sample Environment

Temperature
100 - 300 [K]

Sample Holders

Type
For solid sample these holders are available:
Xyz stepper stage
Xyz+rotation piezo stage
Description
Xyz + rotation around z axis
Detectors

Andor iKon-L SO

Type
Andor iKon-L SO
Passive or Active (Electronics)
Active

Detection

Detected Particle
Photon

HHG-detectors

Type
Spectrometer, X-ray camera, XUV photodiode
Passive or Active (Electronics)
Active

Detection

Detected Particle
Photon

MCPs and Cameras

Passive or Active (Electronics)
Active

Detection

Detected Particle
Photon
contacts
{name}
Techniques
Absorption
  • Time-resolved studies
Diffraction
  • Crystallography
  • Time-resolved studies
Emission or Reflection
  • Ellipsometry
  • Time-resolved studies
Ion Spectroscopy
  • Ion imaging
  • Mass spectrometry
Photoelectron emission
  • Time-resolved studies
Scattering
  • Coherent scattering
  • Elastic scattering
  • Time-resolved scattering
Disciplines
Life Sciences & Biotech
  • Molecular and cellular biology
Physics
  • Atomic & molecular physics
  • Other - Physics
Address
ELI Beamlines
Za radnicí 835
252 41 Dolní Břežany
Czech Republic
control/Data analysis
Control Software Type
  • HHG & MAC: Tango or local
    Users: what they bring
Data Output Format
  • .SPE, .CSV, .himg, .int
Softwares For Data Analysis
  • WinView, oscilloscope, WaveView
Equipment That Can Be Brought By The User